MAGNETOM Scanners recall

MAGNETOM Scanners

Health Canada / Canadian Food Inspection Agency · Nov 5, 2026

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

A software issue has been identified in syngo.via MR Cardiac Analysis MR Flow Quantification affecting the evaluation of retrospectively gated 2D cardiac flow data. - The algorithm does not correctly represent the final phase of the cardiac cycle. The evaluation stops at the last trigger time point (TT point), which marks the beginning of the final cardiac phase, instead of accounting for the complete duration of that phase until the end of the cardiac cycle. The final repetition-time interval (TR) is not included in the calculation ("TR missing"). This results in systematic truncation of the cardiac cycle and incomplete loop closure.- Consequently, flow quantification parameters (e.g., Forward volume, Avg.Flow/Min) are systematically underestimated. - The impact is relevant for Beat Sensor-triggered flow data, for which the flow peak occurs in the very last acquired cardiac phases. The impact is minor or negligible for electrocardiogram (ECG)-triggered flow data, for which the flow values are close to zero in the very last acquired cardiac phases. Contact the manufacturer if you require additional information. Siemens Healthcare Gmbh Henkestr. 127, Erlangen, Germany, 91052

Affected products

  • MAGNETOM Cima.X
    Lots
    Not applicable.
    Models
    11689304, 11647158
  • MAGNETOM Cima.X Fit
    Lots
    Not applicable.
    Models
    11647159
  • MAGNETOM Flow.Elite
    Lots
    Not applicable.
    Models
    11851911, 11878345
  • MAGNETOM Skyra Fit
    Lots
    Not applicable.
    Models
    11516217
  • MAGNETOM Avanto Fit
    Lots
    211077, 211206, 211050, 211139, 211067, 211097
    Models
    11516216

Recall details

Distribution
Canada