Automated Blood Coagulation Analyzers recall

Automated Blood Coagulation Analyzers

Health Canada / Canadian Food Inspection Agency · Nov 6, 2026

Hazard: Insufficient washing of the sample arm B probe independent from the assays used

Check the official notice for affected lots and dates.

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

In October 2026, the legal manufacturer Sysmex notified Siemens Healthineers of an urgent field safety notice for the analyzer CN series systems.Sysmex identified that deposits may accumulate on the sample arm B probe of the CN-series analyzers.The deposits may affect sample dispensing, potentially resulting in the delivery of a lower sample volume to the reaction cuvette than defined, which and may cause falsely decreased results. Based on the manufacturer's investigation, the accumulation of deposits occurred due to insufficient washing of the sample arm B probe independent from the assays used. Contact the manufacturer if you require additional information. Sysmex Corporation 4-4-4- Takatsukadai, Kobe, Japan, 651-2271

Affected products

  • Automated Blood Coagulation Analyzer CN SERIES CN-3000
    Models
    BN343315, CL971029
  • Automated Blood Coagulation Analyzer CN SERIES CN-6000
    Lots
    Q1079, Q1078
    Models
    AX801645, CV339763

Recall details

Distribution
Canada

View original