Automated Blood Coagulation Analyzers recall
Automated Blood Coagulation Analyzers
Hazard: Insufficient washing of the sample arm B probe independent from the assays used
Check the official notice for affected lots and dates.
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
In October 2026, the legal manufacturer Sysmex notified Siemens Healthineers of an urgent field safety notice for the analyzer CN series systems.Sysmex identified that deposits may accumulate on the sample arm B probe of the CN-series analyzers.The deposits may affect sample dispensing, potentially resulting in the delivery of a lower sample volume to the reaction cuvette than defined, which and may cause falsely decreased results. Based on the manufacturer's investigation, the accumulation of deposits occurred due to insufficient washing of the sample arm B probe independent from the assays used. Contact the manufacturer if you require additional information. Sysmex Corporation 4-4-4- Takatsukadai, Kobe, Japan, 651-2271
Affected products
- Automated Blood Coagulation Analyzer CN SERIES CN-3000
- Models
- BN343315, CL971029
- Automated Blood Coagulation Analyzer CN SERIES CN-6000
- Lots
- Q1079, Q1078
- Models
- AX801645, CV339763
Recall details
- Distribution
- Canada