MAGNETOM Flow.Ace recall
MAGNETOM Flow.Ace
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
Siemens Healthineers is advising customers of a software issue affecting syngo.via MR Cardiac Analysis MRFlow Quantification when used with Beat Sensor triggering for retrospectively gated 2D cardiac flowexaminations.The issue affects the MAGNETOM MR scanner model MAGNETOM Flow.Ace when operated with one of thesoftware versions and both licenses specified as follows :. Software Versions Numaris/X VA51A, VA60A, VA61A, VA70A, VB10A, VB20A/NX2510;. Beat Sensor / Pilot Tone Cardiac License (11344791) and. syngo.via MR Cardiac Analysis MR Flow Quantification License (7106227 or 11344838) Contact the manufacturer if you require additional information. Siemens Shenzhen Magnetic Resonance Ltd. Siemens Mri Center, Shenzhen, China, 518057
Affected products
- MAGNETOM Flow.Ace
- Lots
- 218718
- Models
- 11571247
Recall details
- Distribution
- Canada