PD-L1 IHC 22C3 pharmDx recall

PD-L1 IHC 22C3 pharmDx

Health Canada / Canadian Food Inspection Agency · Oct 28, 2026

Hazard: An inadvertent formatting change in section "9.2 tissue sections

Check the official notice for affected lots and dates.

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

Agilent identified missing hyphens in the IFU for PD-L1 IHC 22C3 pharmDx (GE006), revision 02 (PD005650). Due to an inadvertent formatting change in section "9.2 tissue sections," tissue-section thickness was incorrectly listed as 45 μm instead of 4-5 μm, and slide-storage temperature was incorrectly listed as 28 °c instead of 2-8 °c. This issue is limited to the electronic IFU (eIFU) associated with the three lots. Contact the manufacturer if you require additional information. Agilent Technologies, Inc. 5301 Stevens Creek Boulevard, Santa Clara, California, United States, 95051

Affected products

  • Product
    Lots
    12001736B, 12010368B, 11992577B
    Models
    GE006

Recall details

Distribution
Canada