Philips Azurion System recall
Philips Azurion System
Hazard: Energy peaks
Check the official notice for affected lots and dates.
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
Philips has identified a safety issue affecting certain Azurion systems potentially leading to loss of imaging functionality. The power distribution unit is an internal assembly of the Philips Azurion system that contains multiple power supplies. Philips has identified that, in certain Azurion systems, one or more of these power supplies can fail due to energy peaks (surge voltages) during start or restart of the system.When a power supply failure occurs during a start or restart, the power distribution unit does not provide the required direct current power to essential system components. As a result, the system does not successfully complete the (re-)start, imaging functionality is unavailable, and the system remains unavailable for clinical use. Contact the manufacturer if you require additional information. Philips Medical Systems Nederland B.V. Veenpluis 6, Best, Netherlands, 5684 PC
Affected products
- Azurion 7 M20
- Lots
- Not applicable.
- Models
- 722 224, 722 079
- Azurion 7 B12
- Lots
- Not applicable.
- Models
- 722 067
- Azurion 7 M12
- Lots
- Not applicable.
- Models
- 722 223, 722 078
- Azurion 7 B20
- Lots
- Not applicable.
- Models
- 722 226, 722 068
Recall details
- Distribution
- Canada