Siemens Healthineers AG Magnetom and Biograph One Systems recall

Siemens Healthineers AG Magnetom and Biograph One Systems

Health Canada / Canadian Food Inspection Agency · Oct 28, 2026

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

A software issue has been identified in syngo.via MR Cardiac Analysis MR Flow Quantification affecting the evaluation of retrospectively gated 2D cardiac flow data. The algorithm does not correctly represent the final phase of the cardiac cycle. The evaluation stops at the last trigger time point (TT point), which marks the beginning of the final cardiac phase, instead of accounting for the complete duration of that phase until the end of the cardiac cycle. The final repetition-time interval (TR) is not included in the calculation ("TR missing"). This results in systematic truncation of the cardiac cycle and incomplete loop closure. Consequently, flow quantification parameters (e.g., Forward volume, Avg.Flow/Min) are systematically underestimated. The impact is relevant for Beat Sensor-triggered flow data, for which the flow peak occurs in the very last acquired cardiac phases. Contact the manufacturer if you require additional information. Siemens Healthineers Ag Siemensstr. 3, Forchheim, Germany, 91301

Affected products

  • Magnetom Sola
    Models
    11291455
  • Magnetom Sola Fit
    Models
    11410482
  • Magnetom Flow.Elite
    Lots
    234786, 234723, 234827, 234832
    Models
    11851911
  • Magnetom Lumina
    Lots
    196680
    Models
    11344916
  • Biograph One
    Lots
    238015, 238023
    Models
    11689172
  • Magnetom Vida Fit
    Lots
    202203, 202080
    Models
    11410481
  • Magnetom Vida
    Models
    11060815
  • Magnetom Viato.Mobile
    Lots
    239040
    Models
    11689652

Recall details

Distribution
Canada