Horizon Medical Imaging System recall
Horizon Medical Imaging System
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
It was identified that certain replacement monitors for Horizon W/WI systems released into production beginning in January 2025 were shipped without the VGA cable required to connect the monitor to the computer. As a result, field service engineers may have installed the systems using VGA cables from unknown origin. This created a nonconformance related to the use of unapproved components and a potential risk of increased electromagnetic interference (EMI) emissions from VGA cables that have not been evaluated or approved by Hologic. Contact the manufacturer if you require additional information. Hologic, Inc. 600 Technology Drive, Newark, Delaware, United States, 19702
Affected products
- Horizon Medical Imaging System
- Lots
- 309700M, 309701M, 309702M
- Models
- HORIZON WI, HORIZON W
Recall details
- Distribution
- Canada