Durable Medical Equipment (DME) Providers Luna G3 BPAP System (Model LG3800-30VT) recall
Medical Device Correction Update Luna® G3 BPAP System (Model LG3800-30VT)
What to do: Healthcare professionals and DME providers are instructed to discontinue use of Targeted VT on affected devices, inspect and quarantine unused affected products, remove devices from clinical service until corrected...
Check the official notice for affected lots and dates.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
React Health (dba 3B Medical, Inc.) has announced a voluntary medical device correction for the Luna® G3 BPAP System (Model LG3800-30VT) , a Bilevel Positive Airway Pressure (BPAP) System. The recall was originally communicated on October 8, 2026, with an update published on November 17, 2026, and the FDA published the announcement on November 19, 2026. The reason for this correction is a data handling issue where the device, when operated with the Targeted VT feature enabled, may deliver incorrect therapy pressure statistical data. Specifically, the device will not record actual pressure statistical data on its internal SD card nor transmit that data to the cloud-based management system (React Health Connect). This issue affects 7,533 devices distributed within the United States and Puerto Rico. The specific UDI-DI for the affected product is 06948538365731. The affected serial numbers are listed in Appendix A of the official recall notification. The hazardous condition arises only when Targeted VT is enabled and the device is monitored remotely via SD card data or wireless cloud transmission. Inaccurate historical pressure data could lead clinicians to adjust device settings based on incorrect information, potentially delaying therapy until corrections are made. To date, React Health has received fifteen (15) complaints, including two (2) serious adverse reports. The corrective action involves a firmware update to version G3-2.00.80, which React Health will provide to affected customers. Patients using the device are instructed to continue using it as prescribed but must immediately contact their healthcare provider, physician, or DME provider to assess therapy settings and coordinate the firmware correction. They must not independently change prescribed settings or discontinue respiratory therapy without professional guidance. Patients are also required to complete the 'Customer Recall Notification Response Form' from the October 8, 2026 notification and email it to [email protected] within seven (7) days to receive the firmware update SD card. The firmware update will allow accurate pressure data collection and display while Targeted VT is enabled, but historic data collected prior to the update will remain unchanged. Healthcare professionals and DME providers are instructed to discontinue use of Targeted VT on affected devices, inspect and quarantine unused affected products, remove devices from clinical service until corrected, maintain patient therapy with unaffected PAP devices, review historical therapy data with the understanding that it may be inaccurate, notify all downstream recipients of the recall, and coordinate installation of the firmware correction. For assistance, contact React Health Customer Support at 863-226-6285 (Monday through Friday, 8:00 a.m. to 5:00 p.m. Eastern Time) or email [email protected]. Media inquiries should be directed to George Koos at 321-794-3811 or [email protected]. Adverse reactions or quality problems can be reported to the FDA via phone at 1-800-332-1088, online at https://www.accessdata.fda.gov/scripts/medwatch, or by mail/fax using the form from www.fda.gov/MedWatch/getforms.htm.
Affected products
- Luna G3 BPAP System (Model LG3800-30VT)
- UPC
- 06948538365731
- Models
- LG3800-30VT
Recall details
- Firm
- Durable Medical Equipment (DME) Providers, Clinicians
- Distribution
- Puerto Rico, United States
- Contact
- 863-226-6285, 1-800-332-1088, 321-794-3811, 1-888-463-6332