Teleflex has initiated a voluntary recall for specific lots of the above-listed products because there is a potential quality issue affecting the supplied hydrophilic coated spring wire guide (SWG) components. Teleflex has confirmed the coating may swell inconsistently when hydrated with saline leading to clumping of the coating and potential for detachment. Contact the manufacturer if you require additional information. Arrow International LLC Subsidiary Of Teleflex Incorporated 3015 Carrington Mill Blvd., Morrisville, North Carolina, United States, 27560
Affected products
- Spring-Wire Guide
- Models
- PW-18080-HN, PW-18080-HS, PW-16130-HN
- Pressure Injectable PICC Kit
- Models
- PR-35552-HPHNL, PR-35541-HPHNL, PR-35552-HPHNM, PR-35541-HPHNM, PR-35563-HPHNL
- Arrow PICC With Arrowg+ard Blue Advance® Technology, No Placement Wire
- Models
- PR-45552-HPHNL, PR-45552-HPHNM, PR-45563-HPHNL, PR-45541-HPHNM, PR-45563-HPHNM, PR-45541-HPHNL
Recall details
- Distribution
- Canada