Spring-Wire Guide, Pressure Injectable PICC Kit, Arrow PICC With Arrowg+ard Blue recall

Spring-Wire Guide, Pressure Injectable PICC Kit, Arrow PICC With Arrowg+ard Blue Advance® Technology, No Placement Wire

Health Canada / Canadian Food Inspection Agency · Nov 18, 2026

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

Teleflex has initiated a voluntary recall for specific lots of the above-listed products because there is a potential quality issue affecting the supplied hydrophilic coated spring wire guide (SWG) components. Teleflex has confirmed the coating may swell inconsistently when hydrated with saline leading to clumping of the coating and potential for detachment. Contact the manufacturer if you require additional information. Arrow International LLC Subsidiary Of Teleflex Incorporated 3015 Carrington Mill Blvd., Morrisville, North Carolina, United States, 27560

Affected products

  • Spring-Wire Guide
    Models
    PW-18080-HN, PW-18080-HS, PW-16130-HN
  • Pressure Injectable PICC Kit
    Models
    PR-35552-HPHNL, PR-35541-HPHNL, PR-35552-HPHNM, PR-35541-HPHNM, PR-35563-HPHNL
  • Arrow PICC With Arrowg+ard Blue Advance® Technology, No Placement Wire
    Models
    PR-45552-HPHNL, PR-45552-HPHNM, PR-45563-HPHNL, PR-45541-HPHNM, PR-45563-HPHNM, PR-45541-HPHNL

Recall details

Distribution
Canada

View original