Blueprint recall
Blueprint®
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
Stryker has identified the following software bugs with Blueprint® versions 4.2.2 and 4.3.1: ---Version 4.3.1 only: PSI guides generated using the incorrect axis when planning with a perform monopost standard baseplate and a BIO-RSA® Angled graft ---Version 4.2.2 & 4.3.1: Pyrocarbon humeral heads selectable with hrs flex procedures. Pyrocarbon heads are labeled for use with Tornier Flex only. Contact the manufacturer if you require additional information. Tornier S.A.S. 161, Rue Lavoisier, Montbonnot-Saint-Martin, France, 38330
Affected products
- Blueprint®
- Lots
- Version 4.3.1, Version 4.2.2
Recall details
- Distribution
- Canada