DxC 500i Clinical Analyzer and DxC 500i AU Chemistry Module recall

DxC 500i Clinical Analyzer and DxC 500i AU Chemistry Module

Health Canada / Canadian Food Inspection Agency · Nov 18, 2026

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

Beckman Coulter has identified an issue that may affect the reporting of clinical chemistry test results. When a laboratory selects a reporting unit other than the system regional default unit for a clinical chemistry test and multiple tests are ordered under the same sample ID, or multiple replicates of the same test under the same sample ID, the software may incorrectly apply the unit conversion factor more than once. This may result in an erroneous reported value. Contact the manufacturer if you require additional information. Beckman Coulter, Inc. 250 S. Kraemer Blvd., Brea, California, United States, 92821

Affected products

  • DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module
    Lots
    Software 1.5.0 and below
    Models
    C63521, C63522
  • DxC 500 AU Clinical Chemistry Analyzer
    Lots
    Software 1.5.0 and below
    Models
    C63520, C63519

Recall details

Distribution
Canada