Early Alert: Compact Medical Removes and Updates Use Instructions for ButterflyBVM Bag-Valve-Mask Resuscitators

FDA early alert · Source: FDA

What to do: For units that fail or develop a hole during use, they must be labeled "DO NOT USE," quarantined with original packaging, and reported to Compact Medical at [email protected].

Check the official notice for affected lots and dates.

Get free alerts for this notice

Free account. We'll show updates and similar notices in your Dashboard.

The FDA has issued an Early Alert regarding a potentially high-risk issue with the ButterflyBVM Single-Use Bag-Valve-Mask (BVM) Resuscitators manufactured by Compact Medical. The company has issued an "Urgent Medical Device Recall" notice to customers, recommending that certain lots be removed from use or sale, and that other lots undergo updated use instructions (a correction) prior to continued use. The affected product is a manual pulmonary resuscitation and emergency respiratory support device. The issue is the potential for a hole to develop in the bellows of the device, which can cause a tidal volume leak leading to inadequate ventilation and hypoxia, potentially resulting in serious injury or death. As of November 2, 2026, no serious injuries or deaths have been reported. The FDA is continuing to review the issue and will provide updates. The recall action is divided into two categories: removal and correction. For removal, the following lots are affected: ButterflyBVM, READY (SKU R300, Case R308, UDI-DI Each: 00860012073528, Case: 10860012073525) with lot numbers 000109, 000110, 000111, 000112, 000113, 000115, 000116, and 000117; ButterflyBVM (SKU A100, Case A108, UDI-DI Each: 00860012073504, Case: 10860012073501) with lot number 000114. These units should be stopped using and removed as soon as a replacement BVM is in place. For correction, the affected lots are: ButterflyBVM, READY (SKU R300, Case R308) with lot numbers 000081, 000083, 000084, 000086, 000093, 000094, 000097, 000098, 000099, 000100, 000101, 000104, 000107, and 000108; ButterflyBVM (SKU A100, Case A108) with lot numbers 000090, 000091, 000096, and 000105; and ButterflyBVM, PLUS (SKU P200, Case P208, UDI-DI Each: 00860012073511, Case: 10860012073518) with lot numbers 000089, 000092, 000095, and 000106. For devices under correction, users must perform the Pre-Use Test before every use and must not use any unit that fails the test. Backup BVMs should always be available. For units that fail or develop a hole during use, they must be labeled "DO NOT USE," quarantined with original packaging, and reported to Compact Medical at [email protected]. Distributors must stop distributing lots 000109 through 000117, and all personnel should be informed of this recall. The recall notice does not specify particular retail stores, states, or countries of distribution, but the product is a medical device sold in the United States, subject to FDA oversight. The FDA recommends that health care professionals and consumers report any adverse reactions or quality problems to MedWatch, the FDA Safety Information and Adverse Event Reporting Program. The letters to customers were sent on September 26, 2026, and a follow-up was sent on October 7, 2026. All devices from the affected lots, even if packaging is missing, should be assumed to be included. Locations to check include patient care areas, stations, vehicles, crash carts, and storage areas. The product information and instructions for use are available at www.butterflybvm.com/ifu.

Affected products

  • ButterflyBVM, READY
    UPC
    00860012073528, 10860012073525, 00860012073573, 00860012073580
    Lots
    000109, 000110, 000111, 000112, 000113, 000115, 000116, 000117, 000081, 000083, 000084, 000086, 000093, 000094, 000097, 000098, 000099, 000100, 000101, 000104, 000107, 000108
    Models
    R300, R308
  • ButterflyBVM
    UPC
    00860012073504, 10860012073501
    Lots
    000114, 000090, 000091, 000096, 000105
    Models
    A100, A108
  • ButterflyBVM, PLUS
    UPC
    00860012073511, 10860012073518
    Lots
    000089, 000092, 000095, 000106
    Models
    P200, P208