Hospital Glucose Meter Correction: Nova Biomedical Updates Use Instructions for StatStrip
Hospital Glucose Meter Correction: Nova Biomedical Updates Use Instructions for StatStrip Glucose Hospital Meter Systems
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.


The recall concerns the StatStrip Glucose Hospital Meter System manufactured by Nova Biomedical, specifically the most recent "Gen 2.0" model. This device is a hand-held, battery-powered, in vitro diagnostic instrument used with glucose electrochemical test strips to measure glucose in whole blood samples or Quality Control (QC) solutions. The product also stores patient test data, QC data, and operator information. The affected devices include two model numbers with the same unique device identifier (UDI/DI): the StatStrip Glucose Hospital Meter System (Model Number 63685, UDI/DI 00385480636858) and the Glucose Hospital Meter Multipack (Model Number 66739, UDI/DI 00385480636858). The issue applies only to the Gen 2.0 model and does not affect any previously released models. The recall was triggered by reports that when scanning patient ID or Quality Control barcodes, prior information may remain on the meter screen. If accepted, this leads to a glucose result being assigned to the wrong patient's medical record or QC ID. This misassignment could cause healthcare providers to make clinical decisions based on another patient's glucose result, potentially leading to delayed treatment, unnecessary treatment (including inappropriate insulin administration), or failure to treat the intended patient. The FDA has classified this as a Class I recall, the most serious type, as the device may cause serious injury or death if used without correction. As of October 5, Nova Biomedical has reported one serious injury and no deaths associated with this issue. The company has implemented a shipment hold on all new Gen 2.0 meters while corrective actions are developed, with a software update intended for a phased rollout in November 2026. The location data is not specified in the text, but the recall applies to all users of the affected units wherever they may be located, as no specific stores, states, or countries are mentioned in the provided content. The corrective action requires users to follow updated instructions from Nova Biomedical while awaiting the software update. Specifically, users must configure the patient validation setting ("Validate Patient ID's") in the NovaNet system if their Electronic Medical Record (EMR) includes Admission, Discharge, and Transfer (ADT) information, ensuring the checkbox is activated. Additionally, users must configure the meters to automatically log the operator out after each test using the "Logoff" button in the top toolbar and selecting the "Test" mode in the 'Operator Logoff Mode' dropdown. After every barcode scan, users must verify on the device that the patient ID displayed is for the patient currently being tested, confirming the patient name, date of birth, and other identifying information shown on the Confirm Patient ID screen before pressing Accept. If a glucose result populates on the device after testing is complete, users must again verify that the patient ID belongs to the patient currently being tested, not assuming the displayed ID is correct. Customers in the U.S. with adverse reactions, quality problems, or questions about this recall should contact Nova Biomedical at 1(800) 545-6682. The recall does not specify a particular store or retail location for returning the product, as it only applies to hospitals and medical facilities in the U.S. that have purchased the affected device, and the FDA recall notice does not indicate whether the product was sold in other countries.
Affected products
- StatStrip Glucose Hospital Meter System
- Models
- 63685
- Glucose Hospital Meter System Multipack
- Models
- 66739
Recall details
- Contact
- 1(800) 545-6682, 1-888-INFO-FDA (1-888-463-6332)