Free account. We'll show updates and similar recalls in your Dashboard.

The service pack NX VA60A-SP04 is a bug fix update (full installation) that solves/improves the following: Latest it security deliveries and increase in overall stability (syngo MR XA60A-SD12 content included) Installation of multiple SW delta packages now possible at the same time EU DataAact (EY) 2023/2854 implementation:access to all machine data Solved image loading errors for MR Breast Biopsy General MR Protocol Manager improvements Coil specifications adjusted for BM Body 18 Pilot Tone and Spine 32 and 1.5T Sentinelle Breast coils image brightness Supporting IEC 60601-1 compliance, enabling predictable and safe system behavior with regard to table movement Contact the manufacturer if you require additional information. Siemens Healthcare Gmbh Henkestr. 127, Erlangen, Germany, 91052

Affected products

  • Magnetom Prisma
    Lots
    Not applicable.
    Models
    10849582
  • Magnetom Aera
    Lots
    Not applicable.
    Models
    10432914
  • Magnetom Skyra
    Lots
    Not applicable.
    Models
    10432915
  • Manetom Prisma Fit
    Lots
    Not applicable.
    Models
    10849583
  • Magnetom Terra.X
    Lots
    Not applicable.
    Models
    11371478, 11371477