Beaver-Visitec International Mini Blades recall

Beaver-Visitec International Mini Blades

Health Canada / Canadian Food Inspection Agency · Nov 12, 2026

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

BVI identified a potential seal integrity issue affecting certain units of the products listed in this Voluntary Medical Device Recall notice. The issue may result in an "open seal" condition, in which sterility cannot be guaranteed. the condition has been confirmed through physical investigation, including visual inspection and dye penetration testing. Based on the investigation and sampling performed to date, the condition is intermittent and does not affect all units within the affected lots. Contact the manufacturer if you require additional information. Beaver-Visitec International, Inc 500 Totten Pond Rd. - 10 Citypoint, Waltham, Massachusetts, United States, 02451

Affected products

  • Mini Blades
    Lots
    More than 20 lot numbers, contact manufacturer.
    Models
    364823, BEAVER6900, BEAVER6700, BEAVER6400, 364820, 364822

Recall details

Distribution
Canada

View original