MAGNETOM Flow.Elite MR XB10B-HF02 recall
MAGNETOM Flow.Elite MR XB10B-HF02
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
Hotfix syngo MR XB10B-HF02 (NX2510-WSMW) is a bug fix update (delta installation) which includes:• BioMatrix Head/Neck Pro Coil elements detection fixed for the system configurations with Tim[108x24]and Tim[108x48]• Improved data logging to improve error case detection• Improved general system stability during and after start-up• Improved stability and reliability of the patient table Contact the manufacturer if you require additional information. Siemens Healthcare GmbH Henkestr. 127, Erlangen, Germany, 91052
Affected products
- MAGNETOM Flow.Elite
- Lots
- SN 234787, software VB10B-2510, SN 234789, software VB10B-2510
- Models
- 11851911
Recall details
- Distribution
- Canada