Early Alert: Intersurgical Inc. Removes 22mm Smoothbore breathing tubes and systems
Get free alerts for this notice
Free account. We'll show updates and similar notices in your Dashboard.
The FDA has issued an Early Alert regarding a removal of certain Intersurgical 22mm Smoothbore breathing tubes and systems. The action was initiated by Intersurgical Inc., which sent a letter to affected customers on November 3, 2026, recommending the immediate discontinuation of use and quarantine of all stock with affected product/lot numbers. The affected products must be returned to Intersurgical. The FDA is reviewing information about this potentially high-risk device issue and will provide updates as significant new information becomes available. As of November 3, 2026, Intersurgical has reported no serious injuries or deaths associated with this issue. The removal is based on reports from markets outside the U.S. of soft connectors disconnecting from the tube. The investigation found that some products were manufactured with too little Tetrahydrofuran (THF) solvent, which is used to bond the soft connectors onto the Smoothbore tube. If a connector disconnects during use, it could result in a loss or compromise of ventilation, potentially leading to patient desaturation until manual re-initiation, and in severe cases, organ failure, brain damage, or death caused by hypoxia. The affected devices are used to connect patients to machines that help them breathe, carrying air and oxygen to the patient and allowing exhaled air to flow out. The recall affects specific REF numbers and product lot numbers. The affected products and their respective lot numbers are as follows: REF 5018000 – 22mm Smoothbore tube with soft end connectors, 22F-22F, ≥ 1.8m – Lot 32511811 REF 5025000 – 22mm Smoothbore single heated wire breathing system with water trap, ≥ 1.6m – Lot 32522651 REF 5025310 – 22mm Smoothbore single heated wire breathing system with water trap and auto-fill chamber, ≥ 1.6m – Lot 32526115 REF 5026000 – 22mm Smoothbore dual heated wire breathing system, ≥ 1.6m – Lot 32525551 REF 5183000 – 22mm Smoothbore breathing system with monitoring line and exhalation valve, ≥ 1.6m – Lot 32527207 REF 5191000 – 22mm Smoothbore breathing system with monitoring line and exhalation valve, single limb, 1.8m – Lots 32520670, 32522190, 32522802 REF 5191850 – 22mm Smoothbore exhalation valve heated wire breathing system with monitoring line and extra limb, ≥ 1.6m – Lots 32521551, 32521655 REF 5194850 – 22mm Smoothbore exhalation valve heated wire breathing system with extra limb, ≥ 1.6m – Lot 32518127 REF 5804000 – 22mm Smoothbore bilevel breathing system with swivel CO₂ leak port, ≥ 1.8m – Lots 32508034, 32509327, 32510909, 32516895, 32518659, 32519175, 32525543 REF 5805000 – 22mm Smoothbore bilevel breathing system with monitoring line and swivel CO₂ leak port, ≥ 1.6m – Lots 32516315, 32517877 REF 5805030 – 22mm Smoothbore bilevel breathing system with swivel CO₂ leak port and Clear-Guard Midi filter, ≥ 1.8m – Lots 32505469, 32506841, 32513929, 32514276, 32515579, 32518484, 32519342, 32519468, 32519979, 32522143, 32522415, 32523651, 32525498, 32525628, 32526171, 32526270, 32527299, 32528409 The FDA’s notice specifies that customers in the U.S. with adverse reactions, quality problems, or questions should contact Intersurgical Customer Service at [email protected] or 800-828-9633. The text does not provide specific store names, states, or countries beyond noting that the issue was identified from markets outside the U.S. The product removal applies to locations where these devices are used or sold, and the FDA advises health care professionals and consumers to report any adverse reactions or quality problems to MedWatch. The action is based on a risk of ventilation loss, and the FDA will continue to monitor the situation and update its web page as new information becomes available.
Affected products
- 22mm Smoothbore tube with soft end connectors, 22F-22F, ≥ 1.8m
- Lots
- 32511811
- Models
- 5018000
- 22mm Smoothbore single heated wire breathing system with water trap, ≥ 1.6m
- Lots
- 32522651
- Models
- 5025000
- 22mm Smoothbore single heated wire breathing system with water trap and auto-fill chamber, ≥ 1.6m
- Lots
- 32526115
- Models
- 5025310
- 22mm Smoothbore dual heated wire breathing system, ≥ 1.6m
- Lots
- 32525551
- Models
- 5026000
- 22mm Smoothbore breathing system with monitoring line and exhalation valve, ≥ 1.6m
- Lots
- 32527207
- Models
- 5183000
- 22mm Smoothbore breathing system with monitoring line and exhalation valve, single limb, 1.8m
- Lots
- 32520670, 32522190, 32522802
- Models
- 5191000
- 22mm Smoothbore exhalation valve heated wire breathing system with monitoring line and extra limb, ≥ 1.6m
- Lots
- 32521551, 32521655
- Models
- 5191850
- 22mm Smoothbore exhalation valve heated wire breathing system with extra limb, ≥ 1.6m
- Lots
- 32518127
- Models
- 5194850
- 22mm Smoothbore bilevel breathing system with swivel CO₂ leak port, ≥ 1.8m
- Lots
- 32508034, 32509327, 32510909, 32516895, 32518659, 32519175, 32525543
- Models
- 5804000
- 22mm Smoothbore bilevel breathing system with monitoring line and swivel CO₂ leak port, ≥ 1.6m
- Lots
- 32516315, 32517877
- Models
- 5805000
- 22mm Smoothbore bilevel breathing system with swivel CO₂ leak port and Clear-Guard Midi filter, ≥ 1.8m
- Lots
- 32505469, 32506841, 32513929, 32514276, 32515579, 32518484, 32519342, 32519468, 32519979, 32522143, 32522415, 32523651, 32525498, 32525628, 32526171, 32526270, 32527299, 32528409
- Models
- 5805030
Notice details
- Contact
- 800-828-9633