DxC 500i Clinical Analyzer and DxC 500 AU Clinical Chemistry Analyzer recall

DxC 500i Clinical Analyzer and DxC 500 AU Clinical Chemistry Analyzer

Health Canada / Canadian Food Inspection Agency · Oct 26, 2026

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

Beckman Coulter has determined that a modification made by operators to 25 predefined calculated rules, which are included as part of the parameter package, will revert to the default expression following parameter package or software upgrades. Contact the manufacturer if you require additional information. Beckman Coulter, Inc. 250 S. Kraemer Blvd., Brea, California, United States, 92821

Affected products

  • DxC 500 AU Clinical Chemistry Analyzer
    Lots
    1.5 or lower
    Models
    C63520, C63519
  • DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module
    Lots
    1.5 or lower
    Models
    C63521, C63522
  • DxC 500i Clinical Analyzer, DxC 500i Access 2 Immunoassay Module
    Lots
    1.5 or lower
    Models
    C13252

Recall details

Distribution
Canada