TransMedics OCS Liver Perfusion Set recall

OCS Liver Perfusion Set

U.S. Food and Drug Administration · Nov 18, 2026

Hazard: The Perfusion Module may have an unprogrammed Front End Board (FEB) which will not allow the system to proceed beyond the initial startup process.

Check the official notice for affected lots and dates.

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

Class II Recalling firm: TransMedics, Inc. Reason: The Perfusion Module may have an unprogrammed Front End Board (FEB) which will not allow the system to proceed beyond the initial startup process. The OCS system will display an error message saying Unsupported Perfusion Module Present and will prevent further system use. Code: 1. Perfusion Set SN: 4470-4-24374; Description: OCS Liver Perfusion Set; EXP Date: 11/10/27; UDI: +B00944705/$$+74470-4-24374/14D202711100; 2. Perfusion Set SN: 4470-4-23845; Description: OCS Liver Perfusion Set; EXP Date: 10/23/27; UDI: +B00944705/$$+74470-4-23845/14D202710235; 3. Perfusion Set SN: 4470-4-24226; Description: OCS Liver Perfusion Set; EXP Date: 11/10/27; UDI: +B00944705/$$+74470-4-24226/14D20271110$; 4. Perfusion Set SN: 4470-4-24559; Description: OCS Liver Perfusion Set; EXP Date: 11/2/27; UDI: +B00944705/$$+74470-4-24559/14D202711026; 5. Perfusion Set SN: 4470-4-24636; Description: OCS Liver Perfusion Set; EXP Date: 11/2/27; UDI: +B00944705/$$+74470-4-24636/14D202711022; 6. Perfusion Set SN: 4470-4-24508; Description: OCS Liver Perfusion Set; EXP Date: 11/2/27; UDI: +B00944705/$$+74470-4-24508/14D202711020; 7. Perfusion Set SN: 4470-4-24384; Description: OCS Liver Perfusion Set; EXP Date: 11/2/27; UDI: +B00944705/$$+74470-4-24384/14D202711022; 8. Perfusion Set SN: 4470-4-23182; Description: OCS Liver Perfusion Set; EXP Date: 10/23/27; UDI: +B00944705/$$+74470-4-23182/14D20271023%; 9. Perfusion Set SN: 4470-4-23650; Description: OCS Liver Perfusion Set; EXP Date: 10/19/27; UDI: +B00944705/$$+74470-4-23650/14D202710194; 10. Perfusion Set SN: 4470-4-23593; Description: OCS Liver Perfusion Set; EXP Date: 10/19/27; UDI: +B00944705/$$+74470-4-23593/14D20271019A; 11. Perfusion Set SN: 4470-4-23471; Description: OCS Liver Perfusion Set; EXP Date: 10/19/27; UDI: +B00944705/$$+74470-4-23471/14D202710195; 12. Perfusion Set SN: 4470-4-22403; Description: OCS Liver Perfusion Set; EXP Date: 9/10/27; UDI: +B00944705/$$+74470-4-22403/14D20270910+; 13. Perfusion Set SN: 4470-4-22433; Description: OCS Liver Perfusion Set; EXP Date: 9/3/27; UDI: +B00944705/$$+74470-4-22433/14D202709033; 14. Perfusion Set SN: 4470-4-22303; Description: OCS Liver Perfusion Set; EXP Date: 8/20/27; UDI: +B00944705/$$+74470-4-22303/14D20270820/; 15. Perfusion Set SN: 4470-4-22280; Description: OCS Liver Perfusion Set; EXP Date: 8/20/27; UDI: +B00944705/$$+74470-4-22280/14D202708201; Distribution: Worldwide - US Nationwide distribution in the states of CA, FL, IL, MA, MN, NC, NE, NY, TX and the country of Serbia;. Quantity: 18

Affected products

  • OCS Liver Perfusion Set
    UPC
    74470424374, 74470423845, 74470424226, 74470424559, 74470424636, 74470424508, 74470424384, 74470423182, 74470423650, 74470423593, 74470423471, 74470422403, 74470422433, 74470422303, 74470422280
    Models
    3200

Recall details

Recall number
Z-0381-2027
Classification
Class II
Units affected
18
Reason
The Perfusion Module may have an unprogrammed Front End Board (FEB) which will not allow the system to proceed beyond the initial startup process. The OCS system will display an error message saying Unsupported Perfusion Module Present and will prevent further system use.
Recalling firm
TransMedics, Inc.
Distribution
California, Florida, Illinois, Massachusetts, Minnesota, Nebraska, New York, North Carolina, Texas, United States