Vesco Q recall
Vesco Q 100mL Small Bore Bag Feed Set, Model Number VED-059
Hazard: The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%.
Check the official notice for affected lots and dates.
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
Class II Recalling firm: Vesco Medical Llc Reason: The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose, if proper controls are not in place. Over- or under-delivery of enteral nutrition through the pump and associated feed sets could cause temporary or medically reversible adverse health consequences, with a remote probability of serious adverse health consequences. Code: UDI-DI 00810096151377; Lot Numbers: all lots manufactured through 03/26/2026 Distribution: US Nationwide distribution. Quantity: 2,220 units
Affected products
- Vesco Q 100mL Small Bore Bag Feed Set, Model Number VED-059
- UPC
- 00810096151377
Recall details
- Recall number
- Z-0379-2027
- Classification
- Class II
- Units affected
- 2,220 units
- Reason
- The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose, if proper controls are not in place. Over- or under-delivery of enteral nutrition…
- Recalling firm
- Vesco Medical Llc
- Distribution
- Ohio, United States