MAGNETOM Flow.Elite MR XB20A-HF02 recall

MAGNETOM Flow.Elite MR XB20A-HF02

Health Canada / Canadian Food Inspection Agency · Nov 12, 2026

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

Hotfix syngo MR XB20A-HF02 (NX2510-WSMW) is a bug fix update (delta installation) which includes:• BioMatrix Head/Neck Pro Coil elements detection fixed for the system configurations with Tim[108x24]and Tim[108x48]• Improved data logging to improve error case detection• Improved general system stability during and after start-up• Improved stability and reliability of the patient table Contact the manufacturer if you require additional information. Siemens Healthineers AG Siemensstr. 3, Forchheim, Germany, 91301

Affected products

  • MAGNETOM Flow.Elite
    Lots
    SN 234832, Software VB20A-2510, SN 234826, Software VB20A-2510, SN 234825, Software VB20A-2510, SN 234851, Software VB20A-2510, SN 234827, Software VB20A-2510, SN 234818, Software VB20A-2510, SN 234821, Software VB20A-2510
    Models
    11851911, 11878345

Recall details

Distribution
Canada