Vesco Q Enteral Feeding Pump recall
Vesco Q Enteral Feeding Pump, Model Number VED-701, Software Versions 6.0 - 7.8
Hazard: The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%.
Check the official notice for affected lots and dates.
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
Class II Recalling firm: Vesco Medical Llc Reason: The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose, if proper controls are not in place. Over- or under-delivery of enteral nutrition through the pump and associated feed sets could cause temporary or medically reversible adverse health consequences, with a remote probability of serious adverse health consequences. Code: UDI-DI: 00850002550498; Serial Numbers: all serial numbers manufactured though 11/05/2025, Software revisions through 7.8 Distribution: US Nationwide distribution. Quantity: 3,493 units
Affected products
- Vesco Q Enteral Feeding Pump, Model Number VED-701, Software Versions 6.0 - 7.8
- UPC
- 00850002550498
Recall details
- Recall number
- Z-0373-2027
- Classification
- Class II
- Units affected
- 3,493 units
- Reason
- The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose, if proper controls are not in place. Over- or under-delivery of enteral nutrition…
- Recalling firm
- Vesco Medical Llc
- Distribution
- Ohio, United States