Vesco Q Enteral Feeding Pump recall

Vesco Q Enteral Feeding Pump, Model Number VED-701, Software Versions 6.0 - 7.8

U.S. Food and Drug Administration · Nov 18, 2026

Hazard: The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%.

Check the official notice for affected lots and dates.

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

Class II Recalling firm: Vesco Medical Llc Reason: The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose, if proper controls are not in place. Over- or under-delivery of enteral nutrition through the pump and associated feed sets could cause temporary or medically reversible adverse health consequences, with a remote probability of serious adverse health consequences. Code: UDI-DI: 00850002550498; Serial Numbers: all serial numbers manufactured though 11/05/2025, Software revisions through 7.8 Distribution: US Nationwide distribution. Quantity: 3,493 units

Affected products

  • Vesco Q Enteral Feeding Pump, Model Number VED-701, Software Versions 6.0 - 7.8
    UPC
    00850002550498

Recall details

Recall number
Z-0373-2027
Classification
Class II
Units affected
3,493 units
Reason
The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose, if proper controls are not in place. Over- or under-delivery of enteral nutrition…
Recalling firm
Vesco Medical Llc
Distribution
Ohio, United States

View original