Tandem Mobi Insulin Pump with Interoperable Technology (with Android Tandem Mobi recall
Tandem Mobi Insulin Pump with Interoperable Technology (with Android Tandem Mobi Mobile app)
Hazard: A software defect in the Tandem Mobi mobile application for Android (versions 3.5.4, 3.5.2.2, and 3.5.2.1) in which the immediate ( Now ) portion of an Extended Bolus is incorrectly reduced to 0 units when the...
Check the official notice for affected lots and dates.
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
Class II Recalling firm: Tandem Diabetes Care, Inc. Reason: A software defect in the Tandem Mobi mobile application for Android (versions 3.5.4, 3.5.2.2, and 3.5.2.1) in which the immediate ( Now ) portion of an Extended Bolus is incorrectly reduced to 0 units when the entered glucose value equals the user s active target glucose value; the deferred ( Later ) portion is delivered, but the total insulin delivered may be less than intended, and the application does not provide a distinct alert that the immediate portion was omitted, potentially resulting in unrecognized under-delivery of insulin and hyperglycemia. Code: Lot Code: UDI-DI: 00389152547244 Software versions: 3.5.4, 3.5.2.2, and 3.5.2.1 Distribution: Worldwide - US Nationwide and the country of Czech Republic. Quantity: 4086 users
Affected products
- Tandem Mobi Insulin Pump with Interoperable Technology (with Android Tandem Mobi Mobile app)
- Models
- 1015472
Recall details
- Recall number
- Z-0437-2027
- Classification
- Class II
- Units affected
- 4086 users
- Reason
- A software defect in the Tandem Mobi mobile application for Android (versions 3.5.4, 3.5.2.2, and 3.5.2.1) in which the immediate ( Now ) portion of an Extended Bolus is incorrectly reduced to 0 units when the entered glucose value equals the user s active target glucose value; the deferred (…
- Recalling firm
- Tandem Diabetes Care, Inc.
- Distribution
- California, United States