Orthoscan VERSA Accessory Cart recall

Orthoscan VERSA Accessory Cart,

U.S. Food and Drug Administration · Nov 4, 2026

Hazard: The accessory cart weldment does not conform to weld dimensional specifications which may result in separation of the monitor arm mounting structure from the cart.

Check the official notice for affected lots and dates.

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

Class II Recalling firm: ZIEHM-ORTHOSCAN INC Reason: The accessory cart weldment does not conform to weld dimensional specifications which may result in separation of the monitor arm mounting structure from the cart. Code: UDI/ Mini C-Arm serial number (6Axxxx)/Accessory cart serial Number: (VCXXX): 1 6A0110 /VC0110 2 6A0111 /VC0111 3 6A0112 /VC0112 4 6A0115 /VC0113 5 6A0114 /VC0114 6 6A0117 /VC0115 7 6A0116 /VC0116 8 6A0113 /VC0117 9 6A0118 /VC0118 10 6A0119/ VC0119, VC0120; 11 6A0120 /VC0121; 12 6A0121 /VC0122; 13 6A0123 /VC0123; 14 6A0122 /VC0124; 15 6A0126 /VC0125; 16 6A0133 /VC0134; 17 6A0124 /VC0126; 18 6A0128 /VC0128; 19 6A0127 /VC0129; 20 6A0129 /VC0130; 21 6A0125 /VC0127; 22 6A0131 /VC0132; 23 6A0134 /VC0133; 24 6A0142 /VC0144; 25 6A0136 /VC0136; 26 6A0132 /VC0137; 27 6A0137 /VC0138; 28 6A0138 /VC0139; 29 6A0139 /VC0140; 30 6A0140 /VC0141; 31 6A0141 /VC0142; 32 6A0146 /VC0147; 33 6A0143 /VC0143; 34 6A0144 /VC0145; 35 6A0152 /VC0153; 36 6A0147 /VC0148; 37 6A0148 /VC0149; 38 6A0149 /VC0150; 39 6A0150 /VC0151; 40 6A0151 / VC0152; 41 6A0163 / VC0165; 42 6A0130 / VC0131; 43 6A0155 / VC0156; 44 6A0157 / VC0157; 45 6A0156/ VC0158; 46 6A0158 / VC0159, VC0160; 47 6A0159/ VC0161; 48 6A0160/ VC0162; 49 6A0161 / VC0163; 50 6A0162 VC0164; 51 6A0153 VC0154; 52 6A0164 VC0166; 53 6A0165 VC0167 Distribution: Worldwide - US Nationwide distribution in the states of AZ , CA , CO , CT , FL , GA , IL , IN , KS, MA , MD , MI , MN , NC , NE , NV , NY , OH , OR , SC, TX , UT , WA and the countries of Germany , Australia , Spain , Switzerland, Japan. Quantity: 55 units

Affected products

  • Orthoscan VERSA Accessory Cart,
    Models
    1000-0021 and Versa Mini C-Arm System,, 1000-0020

Recall details

Recall number
Z-0286-2027
Classification
Class II
Units affected
55 units
Reason
The accessory cart weldment does not conform to weld dimensional specifications which may result in separation of the monitor arm mounting structure from the cart.
Recalling firm
ZIEHM-ORTHOSCAN INC
Distribution
Arizona, California, Colorado, Connecticut, Florida, Georgia, Germany, Illinois, Indiana, Kansas, Maryland, Massachusetts, Michigan, Minnesota, Nebraska, and 12 more