Instrumentation Laboratory HemoslL AcuStar HIT-lgG(PF-4) recall
HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009802028. in vitro diagnostic (OUS only)
Hazard: Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.
Check the official notice for affected lots and dates.
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
Class I Recalling firm: Instrumentation Laboratory Reason: Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples. Code: Part Number: 0009802028. UDI-DI: 08426950545774. Lot Number (Expiration Date): B40006 (01/16/2027), B38063 (11/01/2026). Distribution: Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, FL, GA, MA, MD, NJ, NY, PA, RI, TN, WA, WV and the countries of Austria, Belgium, Switzerland, Costa Rica, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, India, Italy, Jordan, Japan, South Korea, Luxembourg, Latvia, Netherlands, Saudi Arabia, Sweden, Singapore. Quantity: 1,414 units
Affected products
- HemoslL AcuStar HIT-lgG(PF-4)
- Lots
- 0009802028
Recall details
- Recall number
- Z-0002-2027
- Classification
- Class I
- Units affected
- 1,414 units
- Reason
- Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.
- Recalling firm
- Instrumentation Laboratory
- Distribution
- Arizona, Arkansas, Austria, Belgium, California, Connecticut, Denmark, Finland, Florida, France, Georgia, Germany, Greece, Italy, Latvia, and 15 more