Medtronic MiniMed Generic Name: MiniMed Go App recall
Generic Name: MiniMed Go App Generic Name: MiniMed Go App
Hazard: The missing reminder
Check the official notice for affected lots and dates.
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
Class II Recalling firm: Medtronic MiniMed, Inc. Reason: A software defect in which existing users who had previously configured and enabled the long-acting dose reminder feature in version 1.2.0 or earlier experienced failure of the long-acting dose reminder after updating their app to version 1.2.1. The long-acting dose reminder is an optional feature that, when enabled, triggers a notification at the scheduled time to remind users to take their long-acting insulin dose. The issue was caused by a type cast exception during database migration that occurs upon app update to version 1.2.1 causing a silent failure that prevented the creation of new alerts. Failure to receive the long-acting dose reminder may result in missed or delayed insulin doses, potentially leading to hyperglycemia. Conversely, if users are uncertain whether they took their dose due to the missing reminder, inadvertent duplicate dosing may occur, potentially resulting in hypoglycemia. Code: Lot Code: GTIN for MMT-8600, iOS: 00199150038978 GTIN for MMT-8601, Android: 00199150038985 UDI for MMT-8600, iOS: 00199150038978 UDI for MMT-8601, Android: 00199150038985 Software version 1.2.1 Distribution: Worldwide - U.S. Nationwide distribution in the states of AL, CA, CT, FL, LA, MA, MD, NY, OH, PA, and TX. The countries of Austria, Belgium, Denmark, Finland, Germany, Iceland, Italy, Norway, Spain, Sweden, Switzerland, United Kingdom. Quantity: 623
Affected products
- Generic Name: MiniMed Go App Generic Name: MiniMed Go App
- UPC
- 00199150038978, 00199150038985
- Models
- MMT-8600 (iOS), MMT-8601 (Android)
Recall details
- Recall number
- Z-0265-2027
- Classification
- Class II
- Units affected
- 623
- Reason
- A software defect in which existing users who had previously configured and enabled the long-acting dose reminder feature in version 1.2.0 or earlier experienced failure of the long-acting dose reminder after updating their app to version 1.2.1. The long-acting dose reminder is an optional feature…
- Recalling firm
- Medtronic MiniMed, Inc.
- Distribution
- Alabama, Austria, Belgium, California, Connecticut, Denmark, Finland, Florida, Germany, Iceland, Italy, Louisiana, Maryland, Massachusetts, New York, and 8 more