Corin Medical Unity Knee ASC Modular Handle included in the Corin UNITY ASC recall
Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part Number: 1232040.
Hazard: Potential malfunction of the Modular Handles where the locking mechanism may fail.
Check the official notice for affected lots and dates.
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
Class II Recalling firm: CORIN MEDICAL, LTD. Reason: Potential malfunction of the Modular Handles where the locking mechanism may fail. Any mating instruments already engaged to the handle will not be able to be disengaged, and the handle will be unable to fitted securely to any unengaged mating instruments. Code: Part Number: 1232040. UDI-DI/GTIN; 05056139239656. Lot Code: 097921. Distribution: Worldwide - US Nationwide distribution in the states of CA, GA, IL, MI, NC, RI, TX, UT and the countries of Belgium, France, Germany, Hungary. Quantity: 22 units
Affected products
- Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM
- Lots
- 097921
Recall details
- Recall number
- Z-0356-2027
- Classification
- Class II
- Units affected
- 22 units
- Reason
- Potential malfunction of the Modular Handles where the locking mechanism may fail. Any mating instruments already engaged to the handle will not be able to be disengaged, and the handle will be unable to fitted securely to any unengaged mating instruments.
- Recalling firm
- CORIN MEDICAL, LTD.
- Distribution
- Belgium, California, France, Georgia, Germany, Hungary, Illinois, Michigan, North Carolina, Rhode Island, Texas, United Kingdom, United States, Utah