Corin Medical Unity Knee ASC Modular Handle included in the Corin UNITY ASC recall

Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part Number: 1232040.

U.S. Food and Drug Administration · Nov 11, 2026

Hazard: Potential malfunction of the Modular Handles where the locking mechanism may fail.

Check the official notice for affected lots and dates.

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

Class II Recalling firm: CORIN MEDICAL, LTD. Reason: Potential malfunction of the Modular Handles where the locking mechanism may fail. Any mating instruments already engaged to the handle will not be able to be disengaged, and the handle will be unable to fitted securely to any unengaged mating instruments. Code: Part Number: 1232040. UDI-DI/GTIN; 05056139239656. Lot Code: 097921. Distribution: Worldwide - US Nationwide distribution in the states of CA, GA, IL, MI, NC, RI, TX, UT and the countries of Belgium, France, Germany, Hungary. Quantity: 22 units

Affected products

  • Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM
    Lots
    097921

Recall details

Recall number
Z-0356-2027
Classification
Class II
Units affected
22 units
Reason
Potential malfunction of the Modular Handles where the locking mechanism may fail. Any mating instruments already engaged to the handle will not be able to be disengaged, and the handle will be unable to fitted securely to any unengaged mating instruments.
Recalling firm
CORIN MEDICAL, LTD.
Distribution
Belgium, California, France, Georgia, Germany, Hungary, Illinois, Michigan, North Carolina, Rhode Island, Texas, United Kingdom, United States, Utah

View original