Greiner Bio-One VACUETTE LH Lithium Heparin Separator recall

VACUETTE LH Lithium Heparin Separator

U.S. Food and Drug Administration · Oct 28, 2026

Hazard: Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have...

Check the official notice for affected lots and dates.

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

Class II Recalling firm: Greiner Bio-One North America, Inc. Reason: Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia. Code: Lot Code: Item Number: 454008P Lot Number: B26043FM UDI: Rack: 29120017579032, Case: 39120017579039 Expiration Date: 2027-07-12 Distribution: U.S. Nationwide distribution in the states of CO, FL, IL, MO, NE, NY, OH, PA, and VA. Quantity: 860450

Affected products

  • VACUETTE LH Lithium Heparin Separator
    Lots
    454008P, B26043FM
    Models
    454008P

Recall details

Recall number
Z-0200-2027
Classification
Class II
Units affected
860450
Reason
Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia.
Recalling firm
Greiner Bio-One North America, Inc.
Distribution
Colorado, Florida, Illinois, Missouri, Nebraska, New York, North Carolina, Ohio, Pennsylvania, United States, Virginia

View original