Greiner Bio-One VACUETTE LH Lithium Heparin Separator recall
VACUETTE LH Lithium Heparin Separator
Hazard: Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have...
Check the official notice for affected lots and dates.
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
Class II Recalling firm: Greiner Bio-One North America, Inc. Reason: Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia. Code: Lot Code: Item Number: 454008P Lot Number: B26043FM UDI: Rack: 29120017579032, Case: 39120017579039 Expiration Date: 2027-07-12 Distribution: U.S. Nationwide distribution in the states of CO, FL, IL, MO, NE, NY, OH, PA, and VA. Quantity: 860450
Affected products
- VACUETTE LH Lithium Heparin Separator
- Lots
- 454008P, B26043FM
- Models
- 454008P
Recall details
- Recall number
- Z-0200-2027
- Classification
- Class II
- Units affected
- 860450
- Reason
- Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia.
- Recalling firm
- Greiner Bio-One North America, Inc.
- Distribution
- Colorado, Florida, Illinois, Missouri, Nebraska, New York, North Carolina, Ohio, Pennsylvania, United States, Virginia