Radiometer Medical ApS The ABL90 FLEX and ABL90 FLEX PLUS analyzers are intended recall
The ABL90 FLEX and ABL90 FLEX PLUS analyzers are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting. These tests are only…
Hazard: In vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use and: 1) Sample types, including whole blood samples no longer suitable for testing FHbF and ctBil...
Check the official notice for affected lots and dates.
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
Class II Recalling firm: Radiometer Medical ApS Reason: In vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use and: 1) Sample types, including whole blood samples no longer suitable for testing FHbF and ctBil, 2) Reportable ranges for some parameters, 3) Analyzer configuration, and 4) limitations and warnings. Code: UDI-DI: 05700693930923, 05700693930909 Software versions 3.6 and lower. ABL90 FLEX PLUS Instructions for Use versions prior to: 996-223, Version: 202606O. ABL90 FLEX Instructions for Use versions prior to: 996-189, Version: 202606L Distribution: US Nationwide distribution including in the states of CA, OH, LA, MA, FL, NY, CO, IL, TX, GA, NC, PA, SC, MI, OK, WI, AR, MN, OR, UT, WA, SD, IN, TN, AZ, NE, WY, KY, AL, AK, ID, NJ, DE, MT, CT, MS, MO, VT, WV, IA, MD, KS, DC, NV, VA, NM, NH, ND, ME, HI, RI. Quantity: 5395
Affected products
- The ABL90 FLEX and ABL90 FLEX PLUS analyzers are intended for use by trained technologists, nurses, physicians and therapists
- UPC
- 05700693930923, 05700693930909
Recall details
- Recall number
- Z-0190-2027
- Classification
- Class II
- Units affected
- 5395
- Reason
- In vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use and: 1) Sample types, including whole blood samples no longer suitable for testing FHbF and ctBil, 2) Reportable ranges for some parameters, 3) Analyzer configuration, and 4)…
- Recalling firm
- Radiometer Medical ApS
- Distribution
- Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Denmark, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, and 38 more