Instrumentation Laboratory HemoslL AcuStar HIT-lgG(PF-4) recall

HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009800127. in vitro diagnostic

U.S. Food and Drug Administration · Oct 21, 2026

Hazard: Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.

Check the official notice for affected lots and dates.

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

Class I Recalling firm: Instrumentation Laboratory Reason: Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples. Code: Part Number: 0009800127. UDI-DI: 08426951126545. Lot Number (Expiration Date): B40005 (01/16/2027), B38062 (11/01/2026). Distribution: Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, FL, GA, MA, MD, NJ, NY, PA, RI, TN, WA, WV and the countries of Austria, Belgium, Switzerland, Costa Rica, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, India, Italy, Jordan, Japan, South Korea, Luxembourg, Latvia, Netherlands, Saudi Arabia, Sweden, Singapore. Quantity: 434 units

Affected products

  • HemoslL AcuStar HIT-lgG(PF-4)
    Lots
    0009800127

Recall details

Recall number
Z-0001-2027
Classification
Class I
Units affected
434 units
Reason
Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.
Recalling firm
Instrumentation Laboratory
Distribution
Arizona, Arkansas, Austria, Belgium, California, Connecticut, Denmark, Finland, Florida, France, Georgia, Germany, Greece, Italy, Latvia, and 15 more