QinFlow Warrior Lite Base Unit recall

QinFlow Warrior Lite Base Unit

Health Canada / Canadian Food Inspection Agency · Nov 24, 2026

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

The warrior lite base unit operates in conjunction with a battery and a disposable unit to warm blood and IV fluids prior to administration to patients.Following a technical root-cause investigation into field complaints involving thermal deformation of disposable units, quality in flow determined that an internal mechanical shield was incorrectly assembled on the Printed Circuit Board Assembly (PCBA) during manufacturing. Under harsh environmental conditions or significant mechanical impact, an incorrectly assembled shield may detach from the PCBA and cause an electrical short circuit. Contact the manufacturer if you require additional information. Quality In Flow (QinFlow) Ltd. 23 Ha-Yetsira St., Petah Tikva, Israel, 4951251

Affected products

  • Warrior Lite Base Unit
    Models
    QIF-03

Recall details

Distribution
Canada

View original