Medtronic MiniMed CareLink Clinic recall

CareLink Clinic,

U.S. Food and Drug Administration · Oct 28, 2026

Hazard: CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values entered by healthcare professionals (HCPs) for Total Daily...

Check the official notice for affected lots and dates.

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

Class II Recalling firm: Medtronic MiniMed, Inc. Reason: CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values entered by healthcare professionals (HCPs) for Total Daily Dose (TDD) or Long-acting insulin when calculating the Carb Ratio, Insulin Sensitivity Factor (ISF), and Max Recommended value for Rapid-acting insulin. Instead, the calculator continues to use the system-recommended values, resulting in incorrect rapid-acting insulin parameters. If an HCP or patient inputs these incorrect setting values into the patient's MiniMed Go app, the patient could deliver higher or lower insulin doses compared to their insulin needs, potentially leading to over/under delivery of insulin with a severity ranging from Catastrophic (death from severe hypoglycemia) to Minor (self-treatable hypoglycemia/hyperglycemia). Code: UDI-DI: 0763000B00008748K. Software Versions: 4.4A Software Available: US: Between June 4th, 2026, and June 8th, 2026. Distribution: US Nationwide distribution in the state of KS. Quantity: 1

Affected products

  • CareLink Clinic,
    Models
    MMT-7350

Recall details

Recall number
Z-0203-2027
Classification
Class II
Units affected
1
Reason
CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values entered by healthcare professionals (HCPs) for Total Daily Dose (TDD) or Long-acting insulin when calculating the Carb Ratio, Insulin…
Recalling firm
Medtronic MiniMed, Inc.
Distribution
California, Kansas, United States

View original