RotaFlow Console recall
RotaFlow Console
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
During internal validation of new replacement batteries, MCP identified an issue on the RotaFlow console. This issue can cause the device to stop battery charging after the battery operation mode has been utilized. Further investigation identified a software problem as root cause. Because of this, all devices on the market are considered affected by this issue. Contact the manufacturer if you require additional information. Maquet Cardiopulmonary Gmbh Kehler Strasse 31, Rastatt, Germany, 76437
Affected products
- RotaFlow Console
- Models
- 70104.3292
Recall details
- Distribution
- Canada