Beckman Coulter ROTOR ASSY-SW41, Reference Number 331362, when the rotors are recall

ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839…

U.S. Food and Drug Administration · Oct 28, 2026

Hazard: Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use.

Check the official notice for affected lots and dates.

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

Class II Recalling firm: Beckman Coulter, Inc. Reason: Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result in significant damage to materials (i.e., samples) being processed in the centrifuge and to the centrifuge instrument. Code: Serial Numbers: 24U15250 to 26U15766; Optima UDI-DI: A99846 - 15099590416324, A99842 - 15099590415754, A99839 - 15099590415822, A99836 - 15099590415846, A99833 - 15099590415860 Distribution: US Nationwide distribution. Quantity: 336 units

Affected products

  • ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima…

Recall details

Recall number
Z-0202-2027
Classification
Class II
Units affected
336 units
Reason
Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result in significant damage to materials (i.e., samples) being processed in the centrifuge and to the…
Recalling firm
Beckman Coulter, Inc.
Distribution
Indiana, United States