Baxter Healthcare Mobile column TruSystem 7500 U, Product Code 1730720 recall
Mobile column TruSystem 7500 U, Product Code 1730720
Hazard: The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs.
Check the official notice for affected lots and dates.
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
Class II Recalling firm: Baxter Healthcare Corporation Reason: The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Code: UDI-DI: 00887761974388; Serial number: 110183858 Distribution: US Nationwide distribution in the states of IL, MO, TX. Quantity: 1 unit
Affected products
- Mobile column TruSystem 7500 U, Product Code 1730720
- UPC
- 00887761974388
Recall details
- Recall number
- Z-0333-2027
- Classification
- Class II
- Units affected
- 1 unit
- Reason
- The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse…
- Recalling firm
- Baxter Healthcare Corporation
- Distribution
- Illinois, Missouri, Texas, United States