JenaValve Trilogy Delivery Catheter recall

JenaValve Trilogy Delivery Catheter, Size 27MM

U.S. Food and Drug Administration · Oct 21, 2026

Hazard: Delivery Catheter may have sleeve liner separation that results in difficulty or inability to deploy heart valve device.

Check the official notice for affected lots and dates.

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

Class II Recalling firm: Jenavalve Technology Inc. Reason: Delivery Catheter may have sleeve liner separation that results in difficulty or inability to deploy heart valve device. Code: GTIN - 00850000898158/ Lot: DS-040926-004, DS-041026-008, DS-041026-009, DS-042226-006, DS-042226-007, DS-051526-004, DS-051526-006, DS-060226-007, DS-060526-002, DS-060526-003, DS-060526-007 Distribution: US Nationwide distribution in the states of AZ, CA, FL, GA, IL, MA, MD, MI, MN, MO, NY, OH, PA, SC, TX, UT, VA, WA and the countries of Germany, Denmark, Great Britain, Italy. Quantity: 143 units

Affected products

  • JenaValve Trilogy Delivery Catheter, Size 27MM
    UPC
    00850000898158
    Models
    TRILOGY-DC-L

Recall details

Recall number
Z-0136-2027
Classification
Class II
Units affected
143 units
Reason
Delivery Catheter may have sleeve liner separation that results in difficulty or inability to deploy heart valve device.
Recalling firm
Jenavalve Technology Inc.
Distribution
Arizona, Britain, California, Denmark, Florida, Georgia, Germany, Illinois, Italy, Maryland, Massachusetts, Michigan, Minnesota, Missouri, New York, and 8 more