Think Surgical Square Driver recall
Think Surgical Square Driver
Hazard: Robotic system square drive failed to meet design specifications potentially resulting in pin becoming stuck in square driver or the pin wobbling during drilling.
Check the official notice for affected lots and dates.
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
Class II Recalling firm: Think Surgical, Inc. Reason: Robotic system square drive failed to meet design specifications potentially resulting in pin becoming stuck in square driver or the pin wobbling during drilling. Code: Lot: 25356/Primary DI: M9461091260 Distribution: US Nationwide distribution in the states of NY and DE. Quantity: 15 units
Affected products
- Think Surgical Square Driver
- Lots
- 25356, M9461091260
- Models
- 109126
Recall details
- Recall number
- Z-0192-2027
- Classification
- Class II
- Units affected
- 15 units
- Reason
- Robotic system square drive failed to meet design specifications potentially resulting in pin becoming stuck in square driver or the pin wobbling during drilling.
- Recalling firm
- Think Surgical, Inc.
- Distribution
- California, Delaware, New York, United States