Restor3d Veritas Reverse Total Shoulder Replacement recall
Veritas Reverse Total Shoulder Replacement,
Hazard: (Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components...
Check the official notice for affected lots and dates.
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
Class II Recalling firm: restor3d Inc. Reason: (Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery. Code: All Lot codes/ UDIs:Product Code: 3500-140C3 UDI/GTIN:00840097513492; Product Code: 3500-140U3 UDI/GTIN:00840097513461; Product Code: 3500-140C1 UDI/GTIN:00840097513430; Product Code: 3500-140U1 UDI/GTIN:00840097513409 Distribution: US Nationwide distribution in the states of Texas, Massachusetts, Oklahoma, Florida, Colorado, North Carolina. Quantity: 240 units
Affected products
- Veritas Reverse Total Shoulder Replacement,
- UPC
- 00840097513492, 00840097513461, 00840097513430, 00840097513409
- Models
- 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer Liner
Recall details
- Recall number
- Z-0348-2027
- Classification
- Class II
- Units affected
- 240 units
- Reason
- (Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range…
- Recalling firm
- restor3d Inc.
- Distribution
- Colorado, Florida, Massachusetts, North Carolina, Oklahoma, Texas, United States