Medline surgical convenience kits labeled as: OPEN HEART, Medline Product recall
Medline surgical convenience kits labeled as: OPEN HEART, Medline Product Number/SKU DYNJ902898D
Hazard: Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Check the official notice for affected lots and dates.
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
Class II Recalling firm: Medline Industries, LP Reason: Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use. Code: UDI/DI 10195327369743 (each) 40195327369744 (case). Lot Numbers: 23GBD734, 23EBQ512 Distribution: US Nationwide distribution. Quantity: 165 kits
Affected products
- Medline surgical convenience kits labeled as: OPEN HEART, Medline Product Number/SKU DYNJ902898D
- Lots
- 23GBD734, 23EBQ512
Recall details
- Recall number
- Z-0178-2027
- Classification
- Class II
- Units affected
- 165 kits
- Reason
- Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
- Recalling firm
- Medline Industries, LP
- Distribution
- Illinois, United States