HeartSine Samaritan PAD 350P/360P/500P recall

HeartSine® Samaritan® PAD 350P/360P/500P

Health Canada / Canadian Food Inspection Agency · Nov 25, 2026

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

Stryker has identified a potential issue with the electrical connection between Pad-Pak ™ or Pediatric-Pak™ and the AED. The connection uses spring-loaded electrical contacts, also known as pogo pins. Over time, oxidation of the pogo pin may increase electrical resistance and weaken the internal spring. If the condition progresses, the electrical connection between the Pad-Pak ™ or Pediatric-Pak™ battery and the AED may become intermittent or may be lost. Contact the manufacturer if you require additional information. Heartsine Technologies, Ltd. 207 Airport Road West, Belfast, Northern Ireland, United Kingdom, BT3 9ED

Affected products

  • HeartSine® Samaritan® PAD 350p
    Models
    SAM 350P
  • HeartSine® Samaritan® PAD 500p
    Models
    SAM 500P
  • HeartSine® Samaritan® PAD 360p
    Models
    SAM 360P

View original