Baxter Healthcare TruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating recall
TruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column,
Hazard: There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use.
Check the official notice for affected lots and dates.
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
Class II Recalling firm: Baxter Healthcare Corporation Reason: There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury. Code: Serial numbers: 107971925; UDI-DI number: 00887761974067 Distribution: Worldwide - US Nationwide distribution. Quantity: 1 unit
Affected products
- TruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column,
- UPC
- 00887761974067
- Models
- 1854087
Recall details
- Recall number
- Z-0354-2027
- Classification
- Class II
- Units affected
- 1 unit
- Reason
- There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly,…
- Recalling firm
- Baxter Healthcare Corporation
- Distribution
- Illinois, United States