Howmedica Osteonics TS Universal Baseplate Finishing Tray, Triathlon. recall
TS Universal Baseplate Finishing Tray, Triathlon.
Hazard: Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap.
Check the official notice for affected lots and dates.
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
Class II Recalling firm: Howmedica Osteonics Corp. Reason: Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization. Code: Catalog Number: 6543-6-506. GTIN: 07613327548082. All lot numbers used prior to the correction date of May 8, 2026 are impacted. Distribution: These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS. Quantity: 605 trays
Affected products
- TS Universal Baseplate Finishing Tray, Triathlon.
- UPC
- 65436506, 07613327548082
- Models
- 6543-6-506.
Recall details
- Recall number
- Z-0164-2027
- Classification
- Class II
- Units affected
- 605 trays
- Reason
- Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
- Recalling firm
- Howmedica Osteonics Corp.
- Distribution
- Canada, New Jersey, United States