VascuPuncture PICC High Performance Guidewire recall

VascuPuncture™ PICC High Performance Guidewire

Health Canada / Canadian Food Inspection Agency · Nov 19, 2026

What to do: Users should immediately discontinue use of affected devices, quarantine inventory, and follow recall instructions for return.

Check the official notice for affected lots and dates.

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

Hydrophilic coated guidewire may swell inconsistently when exposed to saline, causing coating clumping and potential particle detachment. Released particles could enter the bloodstream and potentially cause embolic events. Users should immediately discontinue use of affected devices, quarantine inventory, and follow recall instructions for return. Contact the manufacturer if you require additional information. Heraeus Medical Components, LLC 5030 Centerville Road, Saint Paul, Minnesota, United States, 55127

Affected products

  • VascuPuncture™ PICC High Performance Guidewire
    Models
    18070280NW-05MHP, 18145280NW-05MHP

Recall details

Distribution
Canada

View original