Shimadzu Corporation Medical Systems Division Trinias Opera Smart (C16s recall

Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray System containing C-arm support (MH-500, MH-60),

U.S. Food and Drug Administration · Nov 11, 2026

Hazard: The mounting bolts securing the C-arm drive unit of the X-ray system may not be torqued to specification, potentially compromising structural stability and safe operation.

Check the official notice for affected lots and dates.

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

Class II Recalling firm: SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION Reason: The mounting bolts securing the C-arm drive unit of the X-ray system may not be torqued to specification, potentially compromising structural stability and safe operation. Code: Serial number/Shimadzu Medical Systems #/UDI: M1F0D3BF4001/ SMS20250167/ (01)04540217067145(11)250; M1F0CC4EB001/ SMS20250098/ (01)04540217067022(11)241; M1F0D27EA001/ SMS20250001/ (01)04540217067138(11)241; 41F0F43E7001/ SMS20250034/ (01)04540217068951(11)240; M1F0D3BE7001/ SMS20250248/ (01)04540217067145(11)240; M1F0D3BE6001/ SMS20240196/ (01)04540217067145(11)240; 41F0F40E5001/ N/A/ (01)04540217068937(11)240; 41F0F41E5001/ N/A/ (01)04540217068944(11)240; 41F0F43E5001/ SMS20240236/ (01)04540217068951(11)240; M1F0CD7E4001/ SMS20240162/ (01)04540217067046(11)240; 41F0F41E1001/ SMS20240062/ (01)04540217068944(11)240; 41F0F43DB002/ SMS20240101/ (01)04540217068951(11)231; 41F0F43DB001/ SMS20240100/ (01)04540217068951(11)231; M1F0CD7D4001/ SMS20230178/ (01)04540217067046(11)230 Distribution: US Nationwide distribution in the states of SC, TX, FL, TN, CA, FL, NY. Quantity: 118 unts.

Affected products

  • Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray System containing C-arm support (MH-500, MH-60),
    UPC
    04540217067145, 04540217067022, 04540217067138, 04540217068951, 04540217068937, 04540217068944, 04540217067046

Recall details

Recall number
Z-0347-2027
Classification
Class II
Units affected
118 unts.
Reason
The mounting bolts securing the C-arm drive unit of the X-ray system may not be torqued to specification, potentially compromising structural stability and safe operation.
Recalling firm
SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION
Distribution
California, Florida, Japan, New York, South Carolina, Tennessee, Texas, United States