Stryker Leibinger GmbH & Co. KG Precision Pointer Electromagnetic (8000-050-001) recall

Precision Pointer Electromagnetic (8000-050-001)

U.S. Food and Drug Administration · Nov 4, 2026

Hazard: Device may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the device is inside the patient.

Check the official notice for affected lots and dates.

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

Class II Recalling firm: Stryker Leibinger GmbH & Co. KG Reason: Device may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the device is inside the patient. Code: Lot 7000413361 UDI-DI 07613327413205 Distribution: Worldwide - US Nationwide distribution in the states of Alabama, Arizona, California, Colorado, Connecticut, Delaware, Florida, Hawaii, Illinois, Indiana, Kentucky, Louisiana, Massachusetts, Michigan, Missouri, Montana, New Hampshire, New York, Ohio, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Washington and the countries of Austria, Canada, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Norway, Portugal, South Korea, Spain, Switzerland, UK. Quantity: 513

Affected products

  • Precision Pointer Electromagnetic (8000-050-001)
    Lots
    7000413361

Recall details

Recall number
Z-0314-2027
Classification
Class II
Units affected
513
Reason
Device may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the device is inside the patient.
Recalling firm
Stryker Leibinger GmbH & Co. KG
Distribution
Alabama, Arizona, Austria, California, Canada, Colorado, Connecticut, Delaware, Denmark, Finland, Florida, France, Germany, Hawaii, Illinois, and 21 more

View original