Howmedica Osteonics TS Misc. Instruments Tray, Triathlon. recall

TS Misc. Instruments Tray, Triathlon.

U.S. Food and Drug Administration · Oct 21, 2026

Hazard: Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap.

Check the official notice for affected lots and dates.

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

Class II Recalling firm: Howmedica Osteonics Corp. Reason: Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization. Code: Catalog Number: 6543-6-519. GTIN: 07613327555127. All lot numbers used prior to the correction date of May 8, 2026 are impacted. Distribution: These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS. Quantity: 577 trays

Affected products

  • TS Misc. Instruments Tray, Triathlon.
    UPC
    65436519, 07613327555127
    Models
    6543-6-519.

Recall details

Recall number
Z-0167-2027
Classification
Class II
Units affected
577 trays
Reason
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
Recalling firm
Howmedica Osteonics Corp.
Distribution
Canada, New Jersey, United States