Baxter Healthcare Operating table column TS7500 MOBIUS, Hybrid operating table recall

Operating table column TS7500 MOBIUS, Hybrid operating table column,

U.S. Food and Drug Administration · Nov 11, 2026

Hazard: There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use.

Check the official notice for affected lots and dates.

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

Class II Recalling firm: Baxter Healthcare Corporation Reason: There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury. Code: Serial numbers: 108094281, 108468075, 108484417, 108506318, 108516636, 108536785, 108544115, 108604154, 108463223, 108484404, 108489634, 108511411, 108516637, 108544106, 108544116, 108618964, 108463224, 108484405, 108489648, 108511412, 108521436, 108544107, 108604137, 108618965, 108468062, 108484406, 108506316, 108511413, 108521452, 108544110, 108604140, 108468063, 108484408, 108506317, 108516634, 108536776, 108544111, 108604153; UDI-DI number: 00887761974425 Distribution: Worldwide - US Nationwide distribution. Quantity: 77 units

Affected products

  • Operating table column TS7500 MOBIUS, Hybrid operating table column,
    UPC
    00887761974425
    Models
    1704695

Recall details

Recall number
Z-0351-2027
Classification
Class II
Units affected
77 units
Reason
There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly,…
Recalling firm
Baxter Healthcare Corporation
Distribution
Illinois, United States