Intersurgical Inc recall

Breathing Systems Recall: Intersurgical Inc. Removes 22mm Smoothbore Breathing Tubes and Systems

U.S. Food and Drug Administration · Nov 25, 2026

Hazard: A Class I recall

What to do: Intersurgical sent a letter to affected customers on November 3, 2026, instructing them to immediately discontinue use, quarantine any stock with the affected lot numbers, return the product to Intersurgical, share...

Check the official notice for affected lots and dates.

This recall is ongoing.

Get free alerts for this recall

Free account. We'll show updates and similar recalls in your Dashboard.

Intersurgical Inc. has removed multiple 22mm Smoothbore breathing tubes and systems from use and sale due to a Class I recall, the most serious type issued by the FDA. The recall affects devices with specific product/lot numbers, and the FDA has classified this as a potential risk of serious injury or death if use continues. The products are medical devices used to connect patients to ventilators, carrying air and oxygen to the patient and allowing exhaled air to flow out. The issue stems from manufacturing where Tetrahydrofuran (THF) solvent is used to bond soft connectors onto the Smoothbore tube; testing found some products had too little solvent, risking connector detachment and loss of ventilation. As of November 3, 2026, no serious injuries or deaths have been reported. The affected products and their lot numbers are as follows: - REF 5018000: 22mm Smoothbore tube with soft end connectors, 22F-22F, ≥ 1.8m, lot 32511811 - REF 5025000: 22mm Smoothbore single heated wire breathing system with water trap, ≥ 1.6m, lot 32522651 - REF 5025310: 22mm Smoothbore single heated wire breathing system with water trap and auto-fill chamber, ≥ 1.6m, lot 32526115 - REF 5026000: 22mm Smoothbore dual heated wire breathing system, ≥ 1.6m, lot 32525551 - REF 5183000: 22mm Smoothbore breathing system with monitoring line and exhalation valve, ≥ 1.6m, lot 32527207 - REF 5191000: 22mm Smoothbore breathing system with monitoring line and exhalation valve, single limb, 1.8m, lots 32520670, 32522190, 32522802 - REF 5191850: 22mm Smoothbore exhalation valve heated wire breathing system with monitoring line and extra limb, ≥ 1.6m, lots 32521551, 32521655 - REF 5194850: 22mm Smoothbore exhalation valve heated wire breathing system with extra limb, ≥ 1.6m, lot 32518127 - REF 5804000: 22mm Smoothbore bilevel breathing system with swivel CO₂ leak port, ≥ 1.8m, lots 32508034, 32509327, 32510909, 32516895, 32518659, 32519175, 32525543 - REF 5805000: 22mm Smoothbore bilevel breathing system with monitoring line and swivel CO₂ leak port, ≥ 1.6m, lots 32516315, 32517877 - REF 5805030: 22mm Smoothbore bilevel breathing system with swivel CO₂ leak port and Clear-Guard Midi filter, ≥ 1.8m, lots 32505469, 32506841, 32513929, 32514276, 32515579, 32518484, 32519342, 32519468, 32519979, 32522143, 32522415, 32523651, 32525498, 32525628, 32526171, 32526270, 32527299, 32528409 The recall originated from reports outside the U.S. of connectors disconnecting, leading to an internal investigation. The FDA updated this communication on 11/20/2026, classifying the issue as Class I, following an Early Alert on 11/10/2026. Intersurgical sent a letter to affected customers on November 3, 2026, instructing them to immediately discontinue use, quarantine any stock with the affected lot numbers, return the product to Intersurgical, share the notice with all users within the facility, and contact any downstream distributors. The products are used in medical settings, and no specific U.S. states or countries are listed beyond the general distribution, though the issue was reported from markets outside the U.S. Customers in the U.S. with adverse reactions, quality problems, or questions should contact Intersurgical Customer Service at [email protected] or 800-828-9633. Reports of adverse reactions or quality problems can also be submitted to MedWatch, the FDA Safety Information and Adverse Event Reporting Program. The FDA has identified this recall as the most serious type, and continued use may cause serious injury or death. The products are medical devices regulated by the FDA, and the recall involves removal from where they are used or sold. The content is current as of 11/25/2026, and further updates will be communicated through FDA channels.

Affected products

  • 22mm Smoothbore tube with soft end connectors, 22F-22F, ≥ 1.8m
    Lots
    32511811
    Models
    5018000
  • 22mm Smoothbore single heated wire breathing system with water trap, ≥ 1.6m
    Lots
    32522651
    Models
    5025000
  • 22mm Smoothbore single heated wire breathing system with water trap and auto-fill chamber, ≥ 1.6m
    Lots
    32526115
    Models
    5025310
  • 22mm Smoothbore dual heated wire breathing system, ≥ 1.6m
    Lots
    32525551
    Models
    5026000
  • 22mm Smoothbore breathing system with monitoring line and exhalation valve, ≥ 1.6m
    Lots
    32527207
    Models
    5183000
  • 22mm Smoothbore breathing system with monitoring line and exhalation valve, single limb, 1.8m
    Lots
    32520670, 32522190, 32522802
    Models
    5191000
  • 22mm Smoothbore exhalation valve heated wire breathing system with monitoring line and extra limb, ≥ 1.6m
    Lots
    32521551, 32521655
    Models
    5191850
  • 22mm Smoothbore exhalation valve heated wire breathing system with extra limb, ≥ 1.6m
    Lots
    32518127
    Models
    5194850
  • 22mm Smoothbore bilevel breathing system with swivel CO₂ leak port, ≥ 1.8m
    Lots
    32508034, 32509327, 32510909, 32516895, 32518659, 32519175, 32525543
    Models
    5804000
  • 22mm Smoothbore bilevel breathing system with monitoring line and swivel CO₂ leak port, ≥ 1.6m
    Lots
    32516315, 32517877
    Models
    5805000
  • 22mm Smoothbore bilevel breathing system with swivel CO₂ leak port and Clear-Guard Midi filter, ≥ 1.8m
    Lots
    32505469, 32506841, 32513929, 32514276, 32515579, 32518484, 32519342, 32519468, 32519979, 32522143, 32522415, 32523651, 32525498, 32525628, 32526171, 32526270, 32527299, 32528409
    Models
    5805030

Recall details

Contact
800-828-9633, [email protected]

View original