Ventec Life Systems VC+Pro (Japan) recall
VC+Pro (Japan)
Hazard: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Check the official notice for affected lots and dates.
This recall is ongoing.
Get free alerts for this recall
Free account. We'll show updates and similar recalls in your Dashboard.
Class I Recalling firm: Ventec Life Systems, Inc. Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy. Code: All Serial Numbers: UDI: 00855573007211 Distribution: Distribution US nationwide including Puerto Rico. Quantity: 606 units
Affected products
- VC+Pro (Japan)
- UPC
- 00855573007211
- Models
- PRT-00900-100, Package PN:PRT-00973-100. used for ventilatory support.
Recall details
- Recall number
- Z-0125-2027
- Classification
- Class I
- Units affected
- 606 units
- Reason
- Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
- Recalling firm
- Ventec Life Systems, Inc.
- Distribution
- Alabama, Alaska, Arizona, Arkansas, California, Colorado, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, and 33 more